Auditory-Evoked Brainstem Response (ABR) Thresholds
- Equipment and Method for ABR Recording
- Normal ABR Patterns and Threshold
- Elevated ABR Threshold Pattern
- Abnormal ABR Patterns
ABR thresholds in mice are determined using a semi-automated computer system (Intelligent Hearing Systems, Miami, Florida). Subdermal needle electrodes are inserted at the vertex and ventrolaterally to both ears of anesthetized mice. Specific auditory stimuli from 10-100 dB SPL are delivered binaurally through plastic tubes from high frequency transducers. ABR thresholds are obtained, in an acoustic chamber, for clicks and for 8, 16, and 32 kHz pure-tone pips. ABR thresholds of all mice and strains tested are entered in spreadsheet files for storage, easy access, and for the production of periodic progress reports. Click-evoked ABR waveforms, obtained at threshold (T) and at T+10, T+20 and T+30 dB or each mouse, are also stored for future reference. Mice of the CBA/CaJ strain are tested periodically as references for normal hearing, and for monitoring the reliability of the equipment and testing procedures.